ECG-32A /
32A-II Three-Channel ECG Recorder

Features:
- Thermal Array Printing System
- Automatic Measurement of miscellaneous ECG parameters.
- Automatic Baseline Control
- Auto printing of date, time, sensitivity, heart rate, chart speed, filter
status, etc.
- Filter: Adoption of digital filter for AC and EMG.
- Liquid Crystal Display: For main operation menu, AMP menu, recording status
menu, etc.
- Touch-Key Design: With "beep" sound for correct input confirmation.
- Built-in Rechargeable Battery: ensures of at least 50 patients 'ECG
recording'.
- Selectable Output: selectable output mode enables free selection of lead
combination.
- Model ECG32A-II has analysis function.
SPECIFICATIONS(click)
|
DIGITAL THREE CHANNEL ECG-301
|
|
- High Resolution thermal-array Printing System (8
points/mm)
-
Three channels of ECG and annotations including lead mark and such
parameters as sensitivity, paper speed, and filter operation status.
- High Resolution Digital Filter
- One-touch Operation
- Store 12 leads of information and 5 seconds
for each lead.
- Concept of floating input circuit meets IEC
safety standard
- Rechargeable battery supporting more than 50
patient
- Can be updated to six or twelve channels, and
connected to portable computers.
- Real time clock for indicating, archiving and
recalling
- Memory of 12leads ECG signals 5 seconds enables it to
output
|
Specifications(click)
| ECG-1206 six-channel ECG recorder |
ECG-1212 twelve-channel ECG recorder |
 |
<
 |
SPECIFICATIONS(click)
MODEL ECG-6A
Six-channel
ECG recorder
Features:
1.
Large screen clearly displays working status, lead symbol, speed,
sensitivity, and alarm for lead detaching, etc.
2.
3/6-channel record auto mode and manual mode
3.
Digital adjustment baseline and orientation
4.
AC and EMG filter
5.
One-touch button
6.
HR measurement with measure report
7.
Operate on either AC or DC with intellectualized charge system
8.
Recording chart paper 112mm/30m, printing 3/6 channel of ECG wave and
explanation, measurement and analysis report. Optional printing of auto
three-channel + rhythm lead, auto six-channel, manual three-channel +rhythm
lead, etc
9.
Portable, Small volume, and weight only 3.8 kgs (including battery)
Specifications:
1.
Safety: to meet CF standardⅠtype
2.
Input circuit: segregating ground input
3.
Input impedance: ≥50MΩ
4.
Polarization preventing range: ±300mV
5.
Patient leakage current: ≤10μA
6.
Time constant: ≥3.2s
7.
CMRR: ≥100db
8.
Sensitivity control: automatic, 1/2, 1, 2
9.
Frequency response: 0.05-100Hz (-3db)
10.
Sampling rate: 1000d/s
11.
A/D transform: 12 bit
12.
Record mode: 8 Pin/mm (vertical)
13.
Record speed: 25,50mm/s
14.
Power: AC 198-240V, 50-60Hz
DC 12V rechargeable battery
15.
Dimension: 300×249×76mm