GYP-90B INFANT INCUBATOR

 

Air-temp and skin-temp servocontrolled by Micro-computer

With RS232 connector

LED display set-temp, real-temp displays separately

Touch mode Bd

Five systems alarm identification

Isolated over temperature alarm system

Six quiet operate iris ports

Heating power rate0-100%

Skin sensor, air sensor

Equipped with cabinet stand, guardrail, I.V pole;

Humidity meter; Air filter; trmer

Inclination of the bassinet can be adjusted steplessly

 

 TECHNICAL DATA

Power requirement

220V±10%, 50Hz±2% or 110V±10%, 60Hz±2%

Power rate

500VA

Environment temperature

10-30

Relative humidity

30%-75%

Air temp control range

25-37

Skin temp control range

34-37

Air control accuracy

≤±0.5

Skin temp accuracy

≤±0.3

Temperature fluctuation

0.5

Mattress temperature uniformity

0.8

Rise temperature time

45min

Noise level

55dB

Tube power rate

20W

Blue light radiation power rate

500μW/cm

Blue light ray wavelength

420nm-470nm

Timer record range

0-9999hours

Mattress obliquity

0°-8° adjusped stepless