Servocontrolled temperature by Micro-computer,
with RS232 connector, may choose electronic stand according to
requirement, heating power can be adjusted by manual or automatic, set
temp, real-temp and timer display on separate windows, adopted far
infrared quartz heater, X-ray tray under the bassinet, with alarm of
temperature deviation, over temperature sensor failure and power broken,
radiator be adjusted in two ways horizontally, bassinet can be adjusted
steplessly.
Oxygen
inspirator, oxygen cylinder, S-1 low pressure aspirator, HF-1 neonate
resuscitation, TJ-1 infant head fixing, phototherapy light, temperature
sensor, infusion pole and shelves.
|
|
TECHNICAL DATA
|
Voltage
|
220V±10%, 50Hz±2% or 110V±10%, 60Hz±2%
|
|
Power
rate |
≤1000VA(HKN-9010 or 93)
≤1300VA(93B)
≤1400VA(93A)
|
|
Environment temperature
|
20℃-30℃
|
|
Relative humidity
|
30%-75%
|
|
Temp
control range |
25℃-37℃
|
|
Temp
control Accuracy |
±1℃(HKN-93, 93B,93A),
±0.5℃(HKN-9010)
|
|
Measure temp error
|
≤±0.3℃
|
|
Mattress temp uniformily
|
≤2.0℃
|
|
Temperature rise time
|
≤30min
|
|
Far-infrared wavelength
|
1μm~3μm
|
|