Safety

IEC 60601-1 approved, CE marking according to MDD93/42/EEC

Dimension and Weight

 

Dimension: 300mm (W) x 190mm (H) x 305mm (D)

Weight: 8kgs   

Operation Requirements

 

Power: AC100~250V, 50/60 Hz

Temperature: 0~40oC

Humidity: 30~95%  

 

Scope

Neonatal, pediatric, and adult patients

 

Performance Specifications

 

Display: 9" monochrome CRT

Resolution 640x480

4-waveform presentation

Sweep Speed: 6.25mm/s, 12.5mm/s, 5mm/s, 50mm/s

Indicator: Alarm indicator light

Power indicator light

QRS beep and alarm sound

Interface: Parameter module interface

Power input socket

Software upgrade interface (BDM)

External VGA interface for connection of an alternate display

Trend time: 1~72 hours

Alarm: High & Low limits user-adjustable, audible and visual 3-level alarm

Networking: Connected to central monitoring system

Recorder: Built-in, thermal array

Plethysmogram waveform: 2 channels

Record width: 50 mm (report paper 50 mm)

Print speed: 25 mm/s 50 mm/s  

 

ECG

 

5-lead or 3-lead selectable

Input: 5-lead: RA; LA; RL; LL; C

Lead selection: I; II; III; avR; avL; avF; V

ECG waveform: 2 channels

Gain selection: x0.5; x1; x2; x4; auto

Protection: Anti-interference functionality against defibrillator, electrotome, electric grid, and myoelectricity

CMMR: Diagnostic mode: >100dB

Monitoring/operating mode: >120dB

Scaling signal: 1mV±3%

Operation modes: Diagnostic, Monitoring, Operating

Bandwidth: 0.05~130Hz

Alarm range: 15~350bpm

S-T detection: YES

Measurement range: -2.0mV~2.0mV

Alarm range: -2.0mV~2.0mV

Arrhythmia analysis: YES

Pacemaking analysis: YES

Alarm Recall: YES, audible and visual alarm  

Pulse Rate:

 

Range: 25~250bpm

Resolution: 1bpm

Accuracy: ±2bpm

Alarm range: 25-250bpm  

Respiration

 

Method: RA-LL impedance

Measurement range: Adult: 0-120rpm

Child/Neonate: 0-150rpm

Resolution: 1rpm

Accuracy: ±2rpm

Alarm range: Adult:0~120rpm

Child:0~150rpm

Apnea alarm: YES  

NIBP

Method: Automatic oscillometric

Operation modes: Manual/Automatic/Continuous

Measurement unit: mmHg/Kpa selectable

Measurement range: Adult: 15~270mmH

Child: 15~200mmHg

Neonate: 10~135mmHg

Over-pressure protection:Double safety protection

Resolution: 1mmHg

Alarm: Systolic, Diastolic, Mean

Temperature (Surface and Intracavity)

Measurement range: 0~50oC

Resolution: 0.1oC

Accuracy: ± 0.1oC

Channel: Single-channel / Dual- channel

selectable, DT

Alarm range: 0~50oC  

SpO2

Measurement range: 0~100%

Resolution: 1%

Accuracy: ±2% (70~100%)

Alarm range: 0~100%

Pulse Rate:

Range: 25~250bpm

Resolution: 1bpm

Accuracy: ±2bpm

Alarm range: 25-250bpm   

 

Temperature (Surface and Intracavity)

Measurement range: 0~50°C

Resolution: 0.1°C

Accuracy: ± 0.1°C

Channel: Single-channel / Dual- channel

selectable, DT

Alarm range: 0~50°C

 

NOTE: specifications subject to change without prior notice

 