Technical SpecificationsSafetyIEC60601
60601-1 approved, CE marking according to MDD93/42/EEC Dimension and Weight
Dimension: 318mm (W) x 264mm
(H) x 152mm (D)
Weight: 5.4kgs Operation Environment
Power Requirements: AC100-250V,
50/60 Hz
Temperature: 0-40ºC
Humidity: 30-95%
Application Scope:
Adult, pediatric and neonate Performance Specifications
Display: 10.4"/9.4" color TFT
Pixel: 640x480
Language: English
Indicator: 6 waveforms Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Alarm indicator
Power indicator
QRS beep and alarm sound
Interface: Parameter module
cable interface
Power input socket
External VGA interface
Battery:
Built-in and Rechargeable, lasting for maximum 2 hours
Trend time:1-72 hours
Alarm: User-adjustable High and
Low limits 3-level audible and visual alarm
Networking: Connected to
central monitoring system
Recorder:
Built-in, thermal array
Plethysmogram waveform: 2
channel
Record mode: manual, on alarm,
time-defined, etc.
Recording width: 50mm
Recording speed: 25mm/s 50mm/s ECG
5-lead or 3-lead selectable
Input: 5-lead: RA; LA; RL; LL;
C
Lead selection: I; II; III; avR;
avL; avF; V1-V6; CAL
ECG waveform: 2 channels
Gain selection: x0.25; x0.5;
x1; x2; auto
Sweep speed: 12.5mm/s, 25mm/s,
50mm/s
Lead selection: I; II; III; avR;
avL; avF; V1-V6; CAL
ECG waveform: 2 channels
Gain selection: x0.25; x0.5;
x1; x2; auto
Filter modes: Diagnostic,
Monitor, Surgery
Sweep speed: 12.5mm/s, 25mm/s,
50mm/s
Heart Rate range: Adult:
15-300bpm;
Neonate/child: 15-350bpm
Accuracy: ±2bpm or ±2%
Resolution: 1bpm
Current leakage: <10mA
Protection: Withstand
4000VAC/50Hz voltage in isolation. Against electrosurgical interference and
defibrillation
Scaling signal: 1mV±3%
Bandwidth: 0.05-130Hz
Alarm range: Adult: 15-300bpm
Child/neonate: 15-350bpm
S-T segment detection: YES
Measurement range: -2.0mV~2.0mV
Alarm range: -2.0mV~2.0mV
Arrhythmia analysis: YES
Alarm: YES, audible and visual
alarm, alarm events recallable Respiration
Method: RA-LL impedance
Operation modes: Auto/ Manual
Measurement range: Adult:
0-120rpm
Child/neonate: 0-150rpm
Resolution: 1 rpm
Accuracy: ±2 rpm
Alarm range: Adult 0-120 rpm
Child/Neonate 0-150 rpm
Apnea alarm: YES
NIBP
Method: Automatic oscillometric
Operation modes:
Manual/Automatic/Continuous
Measurement unit: mmHg/kPa
selectable
Measurement range: Adult:
15-270mmHg
Child: 15-200mmHg
Neonate: 10-135mmHg
Over-pressure protection:
Double safety protection
Resolution: 1mmHg
Alarm: Systolic, Diastolic,
Mean SpO2
Measurement range: 0-100%
Resolution: 1%
|