YP-930 INFANT INCUBATOR

 

Air-temp or skin-temp servo-controlled by Micro-computer set-temp, real-temp and aux-temp display in three windows, with RS232 connector

Auto-adjust heating power, display heating power 0-100%

Electronical stands

With temperature deviation, over-temp, sensor failure, fan low-speed and power broken alarm and so on

37 temperature set, inclination of the bassinet adjustable steeplessly

Isolated system alarm

Equipped with: guardrail, whole trough (AL), quiet operation windows, double walls hood, air-temp sensor

When opening the front door, automatic air curtain come.

  TECHNICAL DATA

Power requirement

220V±10%, 50HZ±2% or 110V±10%, 60HZ±2%

Power rate

400VA

Environment temperature

20-30

Relative humidity

30%-75%

Air temp control range

25-37 (special set to 39)

Skin temp control range

34-37 (special set to 38)

Air control accuracy

≤±0.5

Skin temp accuracy

≤±0.3

Temperature fluctuation

0.5

Mattress Temperature uniformity

0.8

Rise temperature time

45min

Noise level

55dB

Mattress obliquity

0-8° adjustable stepless