SEMI-AUTOMATIC BIOCHEMICAL ANALYZER MODEL BA-2000III
* perfect
lab managing system under the computer controlling
* print
complete report(
including patient information and the examination result)
* computer can save at least for 2 years biochemical
examination results,
and make statistic works on it.
* no quantity
limit for saving standard curves(
linear or non-linear)
* automatic diagnosis of the course, automatic monitoring
and alarm
* quality controlling system can record and analyze the
results within at least 2 years, print Levy-Jenning quality controlling
chart, X¯, SD, CV%
Specification:
-
light system: quartz halogen lamp 12v / 20w
-
wave length: interference filter: 340nm, 405nm, 492nm, 510nm, 546nm,
578nm, 630nm
-
semi-bandwidth: 8-12nm
-
ABS limit: 0.000-2.500ABS
-
Differentiate rate: 0.001ABS
-
Wave moving: 0.002A / h
-
Colorimeter system: stainless steel / quartz flowing 32ul
-
Temp: room temp, 25ºC,
30ºC,
37ºC
interchangeable, accuracy: 0.1ºC
-
Inter-cross pollution: <2.0%
-
Scattering light: <1% (340nm)
-
Sample volume: 200ul –2000ul adjustable
-
Analysis modes: ABS, end point, continuous monitoring
-
Output: RS232, standard PC
-
Working condition: temp: 15ºC
- 30ºC
Humidity: 85% R.H
- Power: AC220V, 50 / 60Hz
BA-800 SEMI-AUTO BIOCHEMICAL ANALYZER
Features:
l
The operation
system based on windows.
l
Measurement
modes: dynamics, two points, linear end point, non-linear end point,
absorbance, single and double wavelength, sample (reagent) blank.
l
Automatically
print the general report or the report of any item or sample, or print the
reports according to the date, the sample number or the patient number.
l
Quality
control system can show all the statistic results and quality control
charts within 31 days
l
Auto feeding,
printing, storing and adjusting wavelength.
l
Auto
self-diagnosis and indication of problems.
l
Concave
grating is used, better than plane grating. The wavelength is adjustable
continuously. Specifications:
Wavelength
range: 330nm~800nm (continuous)
Wavelength
accuracy: ±1nm
Absorbance
range: -0.3~3.0Abs
Resolution:
0.001Abs
Stability:
<=0.002Abs/hour
Intercross
Pollution: <=1%
Colorimetry
system: light path 10mm, 32ul.
Temp control:
25ºC,
30ºC,
37ºC
or room temp +0.1ºC,
auto control
Sample
Volume: 100-2000ul adjustable
Monitor: LCD
colorful
Printer:
external high resolution printer
Interface:
RS232 dual direction interface
Features:
* The
microprocessor adopts DOC technology
* 16-key
board and mouse for operation
* large LCD display. It will display operating steps and
indicate status
* You just follow the indication to set parameters and
quality controlling samples, each standard curves for items.
* sample examination, record management( save 25000
records)
* print the whole examination report, save and restore the
data even after power–off.
* save 120 items:
I – have saved 40 parameters for common use
II – users can set items or edit parameters or save
permanently.
* display the whole course:
I – diagnosis of abnormal cases such as bladder disturbance
to test again
II – auto modification of abnormal samples, recalculate the
result and no need to measure again.
* complete testing modes: EP, FT, RA, double wave-length,
sample blankness
* real time quality control: I – can set 2 quality
controlling samples for each items
II – can display statistics and
curve of quality control, image and print reports within 30 days
III – RS232 dual direction output
* reliable calibration mode: six standard curves
*
Automatically print the
general report or the report of any item or sample, or print the reports
according to the date, the sample number or the patient number.
Specification:
-
light
system: halogen lamp 6v / 10w
-
wave length: 340nm –700nm. 9 wave
lengths for option:
I – standard: 340nm, 380nm, 405nm, 510nm, 545nm, 578nm,
630nm
II –2 optional between wave length: 450nm, 492nm, 650nm,
690nm III – bandwidth: 8-12nm
-
Absorbance:
2.5ABS
-
Resolution:
I – display: 0.001ABS
II – inter calculation: 0.0001ABS / h
-
accuracy: CV
<=
1%
-
Stability: + 0.003 ABS /
hour after 10 minutes warm-up
-
Inter-cross pollution: <1.5%
-
Cuvette: I –
50ul micro flowing cell
II – temp
control: 25, 30, 37ºC+
0.2ºC
III
-- Sample volume: 200ul –2000ul adjustable
-
monitor:
10.4’’ LCD display for model XD811L
-
16 soft keys
-
External ink
jet printer
- power: 176-264V / 50HZ, 150VA
-
Working
condition: temp: 10ºC
- 35ºC
Humidity: 0-95% R.H |