HA-2000 HEMATOLOGY ANALYZER

Features:

- 3-part differentiation of WBC;17 parameters +3 histogram

- Large screen display for all parameters

- Reagent : diluent lyse and enzymatic cleaner

- Built-in thermal recorder, optional external printer

- Recounting function

- 500 samples' values are stored

- 2 counting mode: venous and capillary

- Sample volume : 40μL (venous blood mode) / 20μL (capillary blood mode)

- Counting speed : 60 samples per hour

- Automatic lyzing, mixing, rinsing and clog-clearing

- Provide concentrated reagent, if required

Technical Specifications

Parameters:

WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, PLT, MPV, PDW, and Histogram for WBC, RBC, PLT

Principles

Electrical resistance for counting and Cyanmethemoglobin for hemoglobin

Performance

Parameter 

Linearity Range Precision (CV%)

WBC(109/L)

 0.0-99.9  3.0%@8.0x109

RBC(1012/L)

 0.00-9.99     2.0%@5.00x1012/L

HGB (g/L)  

 0-240  2.0%@160g/L

MCV (fl)         

40-150  1.0%@90fl

PLT(109/L)

10-999  5.0%@>80x109/L

Sample Volume

Capillary Blood 20ul or 40 ul

Anti-coagulated Venous Blood 40ul

Aperture Diameter

WBC    100 um

RBC/PLT 80 um

Dilution Rations

20ul Capillary Blood WBC 1:512

RBC/PLT               1:45000

Anti-coagulated WBC    1:256

Venous Blood RBC/PLT  1:50625

Speed

60 samples per hour

Reagent Consumption

Diluent 23ml/cycle

Lyse 1.25ml/cycle

E-Z Cleaner 40ml/day

Display

Large 9.4" LCD display (640x480 pixels)

Alerts

19 sets of alert message

Menu

Count, Review, Calibrate, Quality Control, Service, Help

Carryover

Capillary Blood WBC, RBC, PLT < 3%

Anti-coagulated

Venous Blood WBC, RBC, PLT < 3%

Input/Output

RS232x2, 1parallel printer, 1bar code scanner connector (Optional)

Printout

Thermal printing, 8 cm wide paper, various printout formats, external printer optional

Standard Configuration:
Main unit
BC-100 Dilutor
Internal Thermal Recorder
Standard Accessories Kit
Standard Kit of Reagent for Installation