GYP-90b | HKN-9010 | HKN-93 | HKN-93a | HKN-93j | TJY-1D XHZ-90 | YDW-II | YP2000 | YP900 | YP-90a7 | YP930 | YPYP-90A YP-90A Series Infant Incubator
----YP-90A1: The
temp-controller can display real-temp, auxiliary-temp, preset-temp on
different display windows. Six kinds of failure alarms. ----YP-90A3: is based on YP-90A1, adding oxygen density control and display function, four display windows. Can delivery oxygen to baby chamber directly through oxygen connector. The oxygen density monitor can continuously survey the delivery status. Adds alarm of oxygen concentration deviation: ±4%O2 ----YP-90A4: is based on YP-90A1, adding display the incubator's inner humidity. With four display windows. ----YP-90A5: has all functions of YP-90A3, YP-90A4. Five different windows can display set-temp, real-temp, auxiliary-temp, humidity, oxygen density at the same time. ----YP-90A6: is based on YP-90A4, adding humidity control function, equipped with ultrasonic humidity-enhancing controller between 20-85%RH. Adds alarm of humidity deviation: ±8%RH.and water lacking. ----YP-90A7: has all functions of YP-90A3, YP-90A6. The five windows can display set-temp, real-temp, auxiliary-temp, humidity, oxygen concentration. Has nine kinds of self-check failure alarms. ----GYP-90B: Based on YP90A, adding a set of phototherapy unit, and a timer, which can time 99,999 hour continuously.
Features: Temperatures of are controlled by micro-computer automatically. Double- layer hood ensures constant temperature. The inside temperature will remain constant by auto-heating of air curtain when the door open. Quiet Operating window. Self-inspect of failures. The preset temperature and real temperature can be displayed separately on screen. Range of preset temperature: 37-39°C. The bed can be adjusted stepless from 0-8°.
Main technical description: Micro-computer control
Feature: Blue light radiate. The radiate head is rotary at 0° , 30° , 60° . Its stand can be lifted or lowered. It adopts the high-grade electronic ballast. This apparatus can be used with any incubator, rescue table, and infant bed.
It uses the explodeproof micro-crystal quartz far-infrared tube as radiation heat source; Can automatically or manually adjust the heating power(0-100%). The set-temperature and real temperature can be displayed separately. Inclination can be adjusted stepless(±15°C). Radioactive head is rotary (0°,30°,60°)in two directions. The X-ray cassette frame is set under the bed; Alarms of sensor failure, ±2°C deviation, over-temperature and electricity failure. Accessories: low pressure aspirator, APGAR evaluation timer, infant breath resuscitation bag, the phototherapy apparatus to therapy neonate jaundice, infusion stand, two 6.3 L oxygen cylinders made of aluminum alloy, flowmeter, oxygen inhaler.
Based on HKN-93A, adding two sets of Jaundice therapy (white light), but without above accessories. HKN-9010 Infant Warmer (rescue bed) The preset and real temperature can be displayed separately. Far infrared quartzose heating tube. The obliquity of bed and radiating head are adjustable. HKN-90(AP) Neonate Warmer (Rescue Bed) Economic type. The temperature is controlled by the micro-computer. Cold light illumination. Attaches a APGAR timer, which can remind you the caring time. Can control the skin temperature. Far-infrared irradiation. Three kinds of alarms. Bed can be adjusted.
|
||||||||||||||||||||||||||||||||||||||||||||||||
|
Power requirement |
~220V/110V 50Hz/60Hz |
|
Power rate |
≤900W |
|
Temperature Control Range |
25ºC~37ºC |
|
Measure Temp Accuracy |
≤0.3ºC |
|
Mattress Temperature Uniformity |
≤0.8ºC |
|
Environment Temp |
20ºC~30ºC |
|
Far-infrared wavelength |
1μm~3μm |
|
Adjust Angle of Mattress |
±15º
Adjust Steplessly |
|
Adjust Angle of Radiator |
0~90ºC
Adjust in two side |
|
Temperature rise time |
≤35min |
Single-side Phototherapy (blue light); electronic ballast; LED
display timer; lifting mobile pedestal (B-100A has no timer)

Single-side Phototherapy (blue light); electronic ballast; LED
display timer; elegant profile, being put on the top matching with carious
models of infant incubator; (B-200A has no timer)
B-series Infant Phototherapy Unit:
|
SPECIFICATION |
B-100 |
B-200 |
|
Wavelength (blue light) |
≥425-475nm(relative
radiation≥30%) |
|
|
Emittance |
≥550Μw/cm2 |
|
|
Timer Range |
0-999h
59min |
|
|
Timer Accuracy |
2h+1min |
|
|
Angle of light tube (horizontal direction) |
(30º,60º,90º)+5º |
|
|
Scope of light tube |
1100mm-1500mm |
|